HPLC Analysis of Epinephrine and Phenylephrine on Heritage C18 Column According to US Pharmacopeia
Conditions of Experiment
| Column: | Heritage С18 |
| Separation Modes: | reversed-phase |
| Column Dimenstions: | 4.6 x 150 mm, 5 um, 100A |
| Mobile Phase: | Water/MeOH ( 50/50) with 0.1% 1-octanesulfonic acid adjust to pH 3.0 with H3PO4 |
| Detection: | UV 254 nm |
| Sample: | 0.3 mg/ml |
| Injection: | 5 uL |
| Flow rate: | 1 mL/min |
Analytes
| Class of compounds: | Catecholamine, Drug, Neurotransmitter |
| Nature of compounds: | Basic, Hydrophilic, Polar |
| Compounds: | Epinephrine, Phenylephrine |