HPLC Separation of Nucleobases and Nucleosides on Amaze TCH HILIC Mixed-Mode Column
Conditions of Experiment
| Column: | Amaze TCH |
| Separation Modes: | HILIC, cation-exchange |
| Column Dimenstions: | 4.6x150 mm, 3 um, 100A |
| Mobile Phase: | ACN from 92% to 70% in 15 min, 10 mM AmAc pH 5 |
| Detection: | UV 255 nm |
| Sample: | 0.2-0.5 mg/ml |
| Injection: | 2 uL |
| Flow rate: | 1 ml/min |
Analytes
| Class of compounds: | Aromatic base, Nucleobase, Nucleoside |
| Nature of compounds: | Basic, Hydrophilic, Neutral, Polar |
| Compounds: | Thymine, Uracil, Thymidine, Uridine, Adenine, Adenosine, Cytosine, Inosine, Guanine, Cytidine, Guanosine |