Benzenesulfonic acid

HPLC Separation of 15 Organic Acids on Amaze HA Mixed-Mode Column with a Triple Gradient
Conditions of Experiment
Column: Amaze HA
Separation Modes: reveresed-phase , anion-exchange
Column Dimenstions: 4.6x100, 3 um, 100A
Mobile Phase: "A: 2% ACN with 40 mM ammonium phosphate pH 4.0 B: 20% ACN with 100 mM ammonium phosphate pH 2.0 gradient from 100% A to 100% B in 15 minutes, 5 minutes hold "
Detection: 205 nm
Sample: 0.2-0.5 mg/ml
Injection: 2 uL
Flow rate: 1 ml/min
Analytes
Class of compounds: Aromatic acid, Isomer, Organic acid, Organic compound, Preservative, Supplement, Vitamin
Nature of compounds: Acidic, Hydrophilic, Hydrophobic, Neutral, Polar
Compounds: Acetic acid, Ascorbic acid, Formic acid, Succinic acid, Citric acid, Malonic acid, Fumaric acid, Mandelic acid, Oxalic acid, Terephthalic acid, Benzoic acid, Maleic acid, Phthalic acid, Benzenesulfonic acid, p-Toluenesulfonic acid
HPLC Analysis of 6 Aromatic Alcohols, Acids and Neutral Compounds on Amaze Aromatic Stationary Phases.
Application description

Comparison run on Amaze PFP, Amaze AR, Amaze SPF, and Heritage C18 columns.

Conditions of Experiment
Column: Amaze PFP
Separation Modes: reversed-phase, pi-pi
Column Dimenstions: 4.6x150 mm, 5 μm, 100A
Mobile Phase: 50% ACN with 20 mM AmFm pH 3
Detection: UV 255 nm
Sample: 1 mg/ml
Injection: 3 μL
Flow rate: 1 ml/min
Analytes
Class of compounds: Aromatic acid, Aromatic compound, Organic acid, Organic compound, Organic solvent
Nature of compounds: Acidic, Hydrophilic, Hydrophobic, Neutral, Polar
Compounds: Benzyl alcohol, Phenol, Benzoic acid, Toluene, Benzenesulfonic acid
HPLC Separation of Hydrophobic, Hydrophilic, Basic and Acidic Compounds on Amaze TR Mixed-Mode Column
Conditions of Experiment
Column: Amaze TR
Separation Modes: RP, cation-exchange, anion-exchange
Column Dimenstions: 4.6x150 mm, 5 um, 100A
Mobile Phase: 30% ACN with 30 mM AmAc pH 5
Detection: UV 275 nm
Sample: 0.2-0.3 mg/ml
Injection: 3 uL
Flow rate: 1 ml/min
Analytes
Class of compounds: Aromatic base, Aromatic compound, Catecholamine, Drug, Neurotransmitter, Organic acid, Organic compound, Preservative, Stimulant
Nature of compounds: Acidic, Basic, Hydrophilic, Hydrophobic, Neutral, Polar
Compounds: Norphenylephrine, Ephedrine, 3,4-Dihydroxybenzoic acid, Benzenesulfonic acid, Propylparaben, Propranolol
HPLC UV Analysis of Benzenesulfonic, p-Toluenesulfonic and 2-Naphtanesulfonic Acids on Heritage MA Mixed-Mode Column
Conditions of Experiment
Column: Heritage MA
Separation Modes: RP, anion-exchange
Column Dimenstions: 4.6x50 mm, 3 um, 100A
Mobile Phase: ACN with AmFm pH 3
Detection: UV 255 nm
Sample: 0.2 mg/ml
Injection: 3 uL
Flow rate: 1 ml/min
Analytes
Class of compounds: Aromatic acid, Organic acid, Organic compound
Nature of compounds: Acidic, Hydrophilic, Hydrophobic, Neutral, Polar
Compounds: Benzenesulfonic acid, p-Toluenesulfonic acid, 2-Naphthalenesulfonic acid
HPLC Analysis of Benzenesulfonic, p-Toluenesulfonic and 2-Naphtanesulfonic Acids with LC/MS Compatible Conditions on Heritage MA Mixed-Mode Column
Conditions of Experiment
Column: Heritage MA
Separation Modes: RP, anion-exchange
Column Dimenstions: 4.6x50 mm, 3 um, 100A
Mobile Phase: ACN with AmFm pH 3
Detection: UV 255 nm
Sample: 0.2 mg/ml
Injection: 3 uL
Flow rate: 1 ml/min
Analytes
Class of compounds: Aromatic acid, Organic acid, Organic compound
Nature of compounds: Acidic, Hydrophilic, Hydrophobic, Neutral, Polar
Compounds: Benzenesulfonic acid, p-Toluenesulfonic acid, 2-Naphthalenesulfonic acid
HPLC Analysis of Benzenesulfonic and p-Toluenesulfonic Acids on Amaze TR Mixed-Mode Column
Application description

Benzenesulfonic and toluenesulfonic acids are common organic acids used as counter-ions in pharmaceutical industry. Both acids are hydrophilic in nature with strong acidic properties. When traditional RP columns are used, poor retention and peak shape are observed due to non-specific secondary interactions. We developed a robust mixed-mode approach for the separation of benzenesulfonic acid and p-toluenesulfonic acid on the Amaze TR reversed-phase anion- and cation-exchange column. Both acids are retained by a weak reversed-phase mechanism and strong anion-exchange mechanism. The elution of these compounds is controlled by the amount of pH of the buffer, concentration of ions and the amount of acetonitrile in the mobile phase. The method is reproducible and effective with a good retention control and peak shape. It is compatible with major detection techniques including mass spectrometry (MS), Evaporative Light Scattering Detection(ELSD), ultraviolet detection and refractive index detection.

Conditions of Experiment
Column: Amaze TR
Separation Modes: reversed-phase, anion-exchange
Column Dimenstions: 4.6x50 mm, 5 um, 100A
Mobile Phase: 20% ACN with 15 mM AmFm pH 3
Detection: UV 255 nm
Sample: 0.3 mg/ml
Injection: 3 ul
Flow rate: 1 ml/min
Analytes
Class of compounds: Organic acid, Organic compound
Nature of compounds: Acidic, Hydrophilic, Neutral, Polar
Compounds: Benzenesulfonic acid, p-Toluenesulfonic acid