Hydrocortisone

HPLC Analysis of Drug Hydrocortisone on Heritage C18 Column
Conditions of Experiment
Column: Heiratge C18
Separation Modes: reversed-phase
Column Dimenstions: 4.6 x 150 mm, 5 um, 100A
Mobile Phase: 40% ACN with 10 mM NaH2PO4 pH 2.5
Detection: UV 210 nm
Sample: 0.3 mg/ml
Injection: 5 uL
Flow rate: 1 ml/min
Analytes
Class of compounds: Drug, Steroids
Nature of compounds: Hydrophobic, Neutral
Compounds: Hydrocortisone
HPLC Analysis of Drug Hydrocortisone on Heritage C18 Column According to US Pharmacopeia
Conditions of Experiment
Column: Heritage С18
Separation Modes: reversed-phase
Column Dimenstions: 4.6 x 100 mm, 5 um, 100A
Mobile Phase: MeCN/H2O 25/75
Detection: UV 254 nm
Sample: 0.3 mg/ml
Injection: 5 uL
Flow rate: 1 mL/min
Analytes
Class of compounds: Drug, Steroids
Nature of compounds: Hydrophobic, Neutral
Compounds: Hydrocortisone